Assay Method Information

Assay Name:  Automated Patch-Clamp System (OPatch HTX)
Description:  Cells are transferred as suspension in serum-free medium to the QPatch HTX system and kept in the cell storage tank/stirrer during experiments. All solutions applied to cells including the intracellular solution will be maintained at room temperature (19° C. to 30° C.).During the sealing process standard bath solution described above will be used. All solutions applied to cells including the pipette solution will be maintained at room temperature (19° C. to 30° C.). After formation of a Gigaohm seal between the patch electrodes and transfected individual HEK293 cells only Mg-free bath solution will be perfused and the cell membrane will be ruptured to assure electrical access to the cell interior (whole-cell patch-configuration). Inward currents will be measured upon application of 300 μM NMDA (and 8.0 μM Glycine) to patch-clamped cells for 5 sec. During the entire experiment the cells will be voltage-clamped at a holding potential of −80 mV.
Affinity data for this assay
 