Assay Name:  Fluorescence Polarization (FP) Assay
Description:  Fluorescent polarization experiments were performed as follows. Briefly, the fluorescence polarization experiments were read on an Ultra Evolution 384-well plate reader (Tecan) with the 485 nm excitation and 535 nm emission filters. The fluorescence intensities parallel (Intparallel) and perpendicular (Intperpedicular) to the plane of excitation were measured in parallel perpendicular black 384-well NBS assay plates (Corning) at room temperature (20° C). The background fluorescence intensities of blank samples containing the references buffer were subtracted and steady-state fluorescence polarization was calculated using the equation: P=(Intparallel Gintperpendicular)/(Intparallel+Gintperpendicular), and the correction factor G (G=0.998 determined empirically) was introduced to eliminate differences in the transmission of vertically and horizontally polarized light. All fluorescence polarization values were expressed in millipolarization units (mP).